Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2432R (Expiry date: 18.05.2026)

 

Production date : 19.09.2024

Results of quality control 4 (after 20 months), date: 19.05.2026

No serum blank (average) : 0.134

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.004 -0.003 0.001 0.000 0
Negative control serum
(N° 9750-06)
0.030 0.046 0.033 0.039 0.037 2
Weak positive serum
(N° 9750-07)
0.313 0.348 0.322 0.352 0.333 16
Positive control serum
(N° 9750-08)
2.024 2.293 1.999 2.205 2.130 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.130 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 16 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.134 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch