#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2432R (Expiry date: 18.05.2026)
Production date : 19.09.2024
Results of quality control 4 (after
20 months), date: 19.05.2026
No serum blank (average) : 0.134
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.004 |
-0.003 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.030 |
0.046 |
0.033 |
0.039 |
0.037 |
2 |
Weak positive serum
(N° 9750-07) |
0.313 |
0.348 |
0.322 |
0.352 |
0.333 |
16 |
Positive control serum
(N° 9750-08) |
2.024 |
2.293 |
1.999 |
2.205 |
2.130 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9750-08 |
> 1.200 |
2.130 |
Yes |
| Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
16 % |
Yes |
| Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|