#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2532S (Expiry date : 18.07.2027)
Production date : 19.11.2025
Results of quality control 3 (after
12 months), date: 13.11.2026
No serum blank (average) : 0.143
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.003 |
0.004 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.058 |
0.053 |
0.060 |
0.050 |
0.055 |
3 |
Weak positive serum
(N° 9450-07) |
0.745 |
0.701 |
0.639 |
0.723 |
0.702 |
33 |
Positive control serum
(N° 9450-08) |
2.070 |
2.146 |
2.047 |
2.142 |
2.101 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9450-08 |
> 1.200 |
2.101 |
Yes |
| Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
33 % |
Yes |
| Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
3 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.143 |
Yes |
Comments:
Good for use.