Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2533C (Expiry date: 18.07.2027)

Production date : 19.11.2025

Results of quality control 2 (after 6 months), date: 18.05.2026

No serum blank (average) : 0.190

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.007 -0.001 -0.003 -0.004 0.000 0
Negative control serum
(N° 9700-06)
0.090 0.088 0.091 0.081 0.088 4
Weak positive serum
(N° 9700-07)
0.387 0.386 0.353 0.375 0.376 17
Positive control serum
(N° 9700-08)
2.260 2.353 2.210 2.103 2.232 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.232 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.190 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch