Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9200 Toxocara canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis

Lot 2601T (Expiry date: 18.09.2027)

 

Production date : 19.01.2026

Results of quality control 2 (after 6 months), date: 17.07.2026

No serum blank (average) : 0.133

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.003 0.007 -0.003 0.000 0
Negative control serum
(N° 9200-06)
0.012 0.014 0.026 0.012 0.016 1
Weak positive serum
(N° 9200-07)
1.519 1.521 1.531 1.544 1.529 70
Positive control serum
(N° 9200-08)
2.068 2.252 2.229 2.200 2.188 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9200-08 > 1.200 2.188 Yes
Absorbance of weak positive serum 9200-07, as % of positive serum >  32 % 70 % Yes
Absorbance of negative serum 9200-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.133 Yes

Comments:

Good for use.