#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2601T (Expiry date: 18.09.2027)
Production date : 19.01.2026
Results of quality control 2 (after
6 months), date: 17.07.2026
No serum blank (average) : 0.133
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.003 |
0.007 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.012 |
0.014 |
0.026 |
0.012 |
0.016 |
1 |
Weak positive serum
(N° 9200-07) |
1.519 |
1.521 |
1.531 |
1.544 |
1.529 |
70 |
Positive control serum
(N° 9200-08) |
2.068 |
2.252 |
2.229 |
2.200 |
2.188 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9200-08 |
> 1.200 |
2.188 |
Yes |
| Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
70 % |
Yes |
| Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.133 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|