#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2614G (Expiry date: 07.03.2028)
Production date : 08.07.2026
Results of quality control 1 (after
production), date: 14.07.2026
No serum blank (average) : 0.134
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
-0.003 |
0.006 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.145 |
0.123 |
0.156 |
0.146 |
0.143 |
6 |
Weak positive serum
(N° 9350-07) |
1.407 |
1.320 |
1.435 |
1.398 |
1.390 |
58 |
Positive control serum
(N° 9350-08) |
2.331 |
2.452 |
2.425 |
2.420 |
2.407 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9350-08 |
> 1.200 |
2.407 |
Yes |
| Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
58 % |
Yes |
| Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
6 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.