#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2615V (Expiry date: 07.03.2028)

 

Production date : 08.07.2026

Results of quality control 1 (after production), date: 10.07.2026

No serum blank (average) : 0.133

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.002 -0.001 0.001 0.000 0
Negative control serum
(N° 9400-06)
0.099 0.102 0.113 0.115 0.108 5
Weak positive serum
(N° 9400-07)
0.962 0.956 0.951 0.964 0.959 43
Positive control serum
(N° 9400-08)
2.238 2.205 2.215 2.274 2.233 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.233 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 43 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.133 Yes

Comments:

Good for use.