#9310
Echinococcus
multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up
Lot 2625F (Expiry date: 12.06.2028)
Production date : 12.10.2026
Results of quality control 1 (after
production), date: 15.10.2026
No serum blank (average) : 0.113
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
-0.001 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9310-06) |
0.026 |
0.020 |
0.023 |
0.025 |
0.024 |
1 |
Weak positive serum
(N° 9310-07) |
0.157 |
0.149 |
0.161 |
0.146 |
0.154 |
7 |
Positive control serum
(N° 9310-08) |
2.148 |
2.169 |
2.171 |
2.268 |
2.189 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9310-08 |
> 1.200 |
2.189 |
Yes |
| Absorbance of weak positive serum 9310-07, as % of positive serum
|
> 3 %
|
7 % |
Yes |
| Absorbance of negative serum 9310-06, as % of positive serum |
< 3 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Rue
de Morges 22,
CH-1023 Crissier, Switzerland.
Phone: + 41 21 633 31 67. Email: cb@bordier.ch
|