#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2625F (Expiry date: 12.06.2028)

Production date : 12.10.2026

Results of quality control 1 (after production), date: 15.10.2026

No serum blank (average) : 0.113

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.001 -0.001 0.001 0.000 0
Negative control serum
(N° 9310-06)
0.026 0.020 0.023 0.025 0.024 1
Weak positive serum
(N° 9310-07)
0.157 0.149 0.161 0.146 0.154 7
Positive control serum
(N° 9310-08)
2.148 2.169 2.171 2.268 2.189 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.189 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 7 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.113 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Rue de Morges 22, CH-1023 Crissier, Switzerland.
Phone: + 41 21 633 31 67. Email: cb@bordier.ch